Lab manual
lab5-galvo-field
Lab 5 · Galvo field correction
- Goal
- Map and correct in-field distortion, bringing the residual under 15 µm RMS.
- Estimated time
- 60min
- Level
- Intermediate
- Prerequisites
- Complete Labs 1 and 2.
Goal
Map galvo field distortion over a 200 × 200 mm field, build a correction table, and verify the corrected residual.
Steps
Calibration → Galvo field, 9×9 grid, 200 mm field.- Look at the residual heat map: distortion should be small at the centre and largest in the corners (pincushion or barrel).
- Record residual RMS and max before and after correction.
- Re-run with a 5×5 grid. Compare corner residuals — too coarse a grid cannot flatten the corners.
- Re-run with 15×15. See whether the residual keeps falling, and what it costs in acquisition time.
- Use the correction table to mark a 180 mm square and measure the difference between its diagonals.
Pass criteria
| Item | Criterion |
|---|---|
| Residual RMS before correction | typically 80–150 µm |
| Residual RMS after 9×9 correction | < 15 µm |
| Corner max residual after 5×5 | clearly worse than 9×9 |
| Improvement of 15×15 over 9×9 | limited (diminishing returns) |
| Diagonal difference of the square | < 0.1 mm |
Common mistakes
- Grid covering only the centre. The corners are never measured, interpolation becomes extrapolation, and the error explodes.
- Watching RMS but not max. A good RMS hides one corner that is 3× over tolerance — and that corner is where the line will fail.
- Correction table out of step with pixel size. Field correction depends on C1; re-calibrate C1 and you must re-calibrate the field.
- Ignoring jump and mark delays. Wrong values leave tails on marked lines that are easily confused with field distortion.
Review questions
- Where does field distortion come from? (Hint: residual design error in the f-θ lens, and the two mirrors not being co-planar.)
- Going from 9×9 to 15×15 multiplies acquisition by 2.8× for a small gain. What metric tells you where to stop?
- How does the residual change if the field grows from 200 mm to 400 mm? Why are large fields harder?
- Bilinear interpolation versus a 2D polynomial fit for the correction table: what does each buy you, and which is more robust to acquisition noise?